Model Number S404
Medical Device Problem Codes Battery Problem (2885); Impedance Problem (2950)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/01/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that this pacemakers longevity in november was 1.5 years and then 6 months later it was at elective replacement indicator (eri).Additionally, it was noted they were unable to obtain impedance measurements however, daily measurements were present.Amplitudes and thresholds were within normal range.At this time, the pacemaker remains in service.No adverse patient effects were reported.
 
Event or Problem Description
It was reported that this pacemakers longevity in november was 1.5 years and then 6 months later it was at elective replacement indicator (eri).Additionally, it was noted they were unable to obtain impedance measurements however, daily measurements were present.Amplitudes and thresholds were within normal range.At this time, the pacemaker remains in service.No adverse patient effects were reported.
 
Additional Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, detailed mechanical and electrical testing was performed on the device.The device battery status was eri.The device functioned normally throughout testing.Factors influencing the estimated longevity remaining calculation include pacing rate, amplitude, pulse-width and lead impedance.Any (even slight) changes in these factors will impact the battery consumption calculation and therefore the remaining longevity estimate.Please note that, despite the drop in estimated longevity remaining, the actual battery condition did not change significantly between follow-ups.In summary, it was determined that this device experienced normal battery depletion but declared eri earlier than previously estimated.Of note, the estimated remaining battery longevity of this pacemaker was designed to be calculated (and trigger corresponding device replacement indicators) based on the pacing capacitor charge time, which progressively increases as the battery depletes.These pacemakers were not designed with an ability to calculate remaining longevity for every possible current drain, which may cause replacement indicators to trigger earlier than what was previously estimated; however, it does not negatively impact the use, operation, safety, or performance of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALTRUA 40
Common Device Name
IMPLANTABLE DEVICE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key14622700
Report Number2124215-2022-20109
Device Sequence Number19947391
Product Code LWP
UDI-Device Identifier00802526484452
UDI-Public00802526484452
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
N970003/S106
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/27/2011
Device Model NumberS404
Device Catalogue NumberS404
Device Lot Number591573
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/10/2022
Supplement Date Received by Manufacturer04/05/2023
Initial Report FDA Received Date06/07/2022
Supplement Report FDA Received Date04/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/04/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-