It was reported that this pacemakers longevity in november was 1.5 years and then 6 months later it was at elective replacement indicator (eri).Additionally, it was noted they were unable to obtain impedance measurements however, daily measurements were present.Amplitudes and thresholds were within normal range.At this time, the pacemaker remains in service.No adverse patient effects were reported.
|
|
It was reported that this pacemakers longevity in november was 1.5 years and then 6 months later it was at elective replacement indicator (eri).Additionally, it was noted they were unable to obtain impedance measurements however, daily measurements were present.Amplitudes and thresholds were within normal range.At this time, the pacemaker remains in service.No adverse patient effects were reported.
|
|
Upon receipt at our post market quality assurance laboratory, detailed mechanical and electrical testing was performed on the device.The device battery status was eri.The device functioned normally throughout testing.Factors influencing the estimated longevity remaining calculation include pacing rate, amplitude, pulse-width and lead impedance.Any (even slight) changes in these factors will impact the battery consumption calculation and therefore the remaining longevity estimate.Please note that, despite the drop in estimated longevity remaining, the actual battery condition did not change significantly between follow-ups.In summary, it was determined that this device experienced normal battery depletion but declared eri earlier than previously estimated.Of note, the estimated remaining battery longevity of this pacemaker was designed to be calculated (and trigger corresponding device replacement indicators) based on the pacing capacitor charge time, which progressively increases as the battery depletes.These pacemakers were not designed with an ability to calculate remaining longevity for every possible current drain, which may cause replacement indicators to trigger earlier than what was previously estimated; however, it does not negatively impact the use, operation, safety, or performance of the device.
|