| Catalog Number |
600-9007-25 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/25/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2022, abbott point of care was contacted by a customer regarding i-stat ec8+ cartridges that yielded a suspected discrepant sodium result of 129 mmol/l on a patient.There was no patient information available at the time of this report.Method, date, result(na); i-stat, (b)(6) 2022, 129 mmol/l; lab, (b)(6) 2022, 150 mmol/l; test/collection times not provided.Cartridge lot# information is also unknown at the time of this report.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on the limited information available that suggests the product was not performing within the variability of the assay.
|
| |
|
Additional Manufacturer Narrative
|
|
Apoc incident: # (b)(4).The investigation was completed on (b)(6) 2022.The customer reported obtaining unexpected na (sodium) patient sample results while using i-stat ec8+ cartridge lot unknown.A review of na performance verification data of i-stat versus multiple reference instruments at multiple sites for clew a42 and a43 demonstrates that the overall performance of the i-stat na is in agreement with other manufacturers.No deficiency has been determined.
|
| |
|
Event or Problem Description
|
|
Na.
|
| |
|