• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL RIGIDLOOP CORTICAL FIXATION IMPLANT 20MM; SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL RIGIDLOOP CORTICAL FIXATION IMPLANT 20MM; SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE Back to Search Results
Model Number 232011
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/20/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).Reporter is a j&j sales representative.Investigation summary: the complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.A manufacturing record evaluation was performed for the finished device lot number (8l58459), and no non-conformance was identified.As part of depuy mitek's quality process all devices are manufactured, inspected, and released to approved specifications.With the information provided, and without the complaint device to evaluate, we cannot determine a root cause for the reported failure.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Event or Problem Description
It was reported by sales rep that during an unknown procedure on (b)(6) 2022, it was observed that the suture on the rigidloop cortical fixation implant 20mm device was frayed.Another like device was used to complete the procedure.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RIGIDLOOP CORTICAL FIXATION IMPLANT 20MM
Common Device Name
SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin blanc 38
le locle CH-24 00
SZ  CH-2400
Manufacturer (Section G)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle CH-24 00
SZ   CH-2400
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key14626188
Report Number1221934-2022-01669
Device Sequence Number4637931
Product Code MBI
UDI-Device Identifier10886705022656
UDI-Public10886705022656
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K130814
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2024
Device Model Number232011
Device Catalogue Number232011
Device Lot Number8L58459
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/20/2022
Initial Report FDA Received Date06/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-