• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES CORPORATION OH/USA ULTRA 360 PATIENT POSITIONING SYSTEM; N/A

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES CORPORATION OH/USA ULTRA 360 PATIENT POSITIONING SYSTEM; N/A Back to Search Results
Model Number A2009
Medical Device Problem Code Device Slipped (1584)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
This is 1 of 3 reports linked to mfg report numbers: 3004608878-2022-00118, 3004608878-2022-00121.A facility reported that the mayfield ultra 360 patient positioning system (a2009) slipped when patient's head was secured in it.No patient injury or surgical delay has been reported.
 
Additional Manufacturer Narrative
The ultra 360 patient positioning system (a2009) was returned for evaluation: device history record (dhr): the dhr was reviewed and shows no abnormalities related to the reported failure.The investigation of the returned device showed that the reported complaint was confirmed from the evaluation.Unit received had the upper and lower handles adjusted to bring them back into tolerance from not holding properly.Both shock cushions, ¼ inch pin, and adjustment wrench were replaced from being worn.Unit also received with the standard adaptor and not the tri-star adaptor.Root cause - complaint confirmed via inspection of the unit.Both the upper and lower handles were not holding properly and required readjustment.Additionally, the unit required replacement of worn components.Root cause is routine use and wear of the device.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRA 360 PATIENT POSITIONING SYSTEM
Common Device Name
N/A
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key14627727
Report Number3004608878-2022-00117
Device Sequence Number12031400
Product Code FWZ
UDI-Device Identifier10381780267850
UDI-Public10381780267850
Combination Product (Y/N)N
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberA2009
Device Catalogue NumberA2009
Device Lot Number134
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/06/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/17/2022
Initial Report FDA Received Date06/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/01/2013
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-