| Model Number |
0186 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/04/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This investigation will be updated should further pertinent information be provided.
|
| |
|
Event or Problem Description
|
|
It was reported that this right ventricular (rv) lead has been showing high, out of range shock impedances that have been increasing over time.More analysis was recommended but the rv lead remains in service at this point.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This investigation will be updated should further pertinent information be provided.
|
| |
|
Event or Problem Description
|
|
It was reported that this right ventricular (rv) lead has been showing high, out of range shock impedances that have been increasing over time.More analysis was recommended.At this time the rv lead has been explanted and returned for analysis.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the lead was performed.Testing was completed to assess lead electrical performance.Measurements were within normal limits.Microscopic inspections of the terminal pin assembly and the returned part of the lead body found no anomalies.Laboratory testing was unable to reproduce the reported clinical observations and detailed analysis did not reveal any abnormalities.Laboratory analysis did not identify any lead characteristics that would have caused or contributed to the reported clinical observations.This investigation will be updated should further pertinent information be provided.
|
| |
|
Event or Problem Description
|
|
It was reported that this right ventricular (rv) lead has been showing high, out of range shock impedances that have been increasing over time.More analysis was recommended.At this time the rv lead has been explanted and returned for analysis.No additional adverse patient effects were reported.
|
| |
|