H3, h6: the product has not been received at the lyon site for evaluation and photographs were not provided, so the complaint could not be confirmed.The following investigative actions were performed.Refer to the respective tasks for further detail.Complaint history review: the complaint history review identify three previous similar reported event for this device (trend on manufacturing lot cannot be done as no lot number was provided in the reported information).No adverse trend was identified.Future complaints for this failure mode will continue to be monitored and investigated as required.Risk management review (device): identified no issues which could have caused or contributed to the reported event.The failure mode was previously identified by the risk management file and the anticipated risk level is acceptable.Dhr/batch record review: no review of the dhr can be done as not manufacturing lot was provided in the reported information.Capa/nc/pra/hhe/field action review: identified no previous events or issues which could have caused or contributed to the reported event.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.Device labeling/ifu review: identified no issues which could have caused or contributed to the reported event.This review determined that the device met labeling requirements upon release for distribution.Medical investigation monitoring board (mimb) review: per mimb, a medical investigation was not required.The complaint alleges no injury to the patient.It could not be determined whether the device contributed to the reported event.According to risk management documentation for the hallulock instruments, application of excessive mechanical forces, improper check prior to the use or improper raw material are identified as failure mode which could result in breakage of the hallu lock driver.Based on this investigation, the need for corrective action is not indicated as no non-conformances or product deficiencies were identified and the risk level is acceptable.If additional information is later received, the complaint may be reopened.No further investigation is warranted for this complaint, though future complaints will be monitored and investigated as required.This investigation can be closed.Internal complaint reference number: case-(b)(4).
|