| Model Number |
MS1-4590R |
| Medical Device Problem Codes |
Corroded (1131); Fracture (1260); Naturally Worn (2988); Scratched Material (3020)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
04/20/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
No product has been returned for evaluation.Evaluation was performed by london implant retrieval centre (lirc).The root cause is unable to be determined at this time.If any additional information is provided, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
An investigation report from an implant retrieval center was received and reported that a revision procedure was performed due to the rod experiencing a pin fracture.During examination of the rod, the retrieval center found corrosion and discoloration on the rod.Additionally, wear, scratches, and other surface damage was noted.There was no further reported patient impact and no patient or user injury was reported.No additional information is available.Report 2 of 2.
|
| |
|
Event or Problem Description
|
|
No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional data: b5, g3, g6, h2, h6, h10.Report 2 of 2.A review of the dhr documents indicated that the rod was manufactured by the specified requirements and met all of the required quality inspections prior to shipment.
|
| |
|