• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM - MAGNETIC ACTUATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM - MAGNETIC ACTUATION Back to Search Results
Model Number MC2-5090S
Medical Device Problem Codes Mechanical Problem (1384); Naturally Worn (2988); Scratched Material (3020)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 07/08/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device has not been returned for evaluation.The device evaluation was performed by the london implant retrieval center (lirc).The root cause is unable to be determined at this time.If any additional information is provided, a supplemental report will be submitted.
 
Event or Problem Description
An investigation report from an implant retrieval center was received and reported that a revision procedure was performed due to the rod failing to distract.During examination of the rod, the retrieval center found the rod unable to distract.Additionally, discoloration, wear, scratches, and other surface damage was noted.There was no further reported patient impact and no patient or user injury was reported.No additional information is available.Report 1 of 2.
 
Event or Problem Description
No additional information has been provided.
 
Additional Manufacturer Narrative
Corrected data: d4.Additional data: h6, h10.A review of the device history records (dhr) confirmed the device met all quality inspections and specifications prior to release.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Common Device Name
GROWING ROD SYSTEM - MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr ste 100
aliso viejo CA 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr ste 100
aliso viejo CA 92656
Manufacturer Contact
geoff gannon
101 enterprise dr ste 100
aliso viejo, CA 92656
8583448112
MDR Report Key14629745
Report Number3006179046-2022-00198
Device Sequence Number6169505
Product Code PGN
UDI-Device Identifier00887517834041
UDI-Public887517834041
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K171791
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Other,Foreign
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 06/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMC2-5090S
Device Lot Number7091601AAA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/10/2022
Supplement Date Received by Manufacturer06/27/2022
Initial Report FDA Received Date06/08/2022
Supplement Report FDA Received Date06/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/16/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age10 YR
Patient SexFemale
Patient Weight39 KG
-
-