Model Number ATG120164
Medical Device Problem Codes Material Rupture (1546); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/17/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history record is currently being performed.The return of the sample is pending.However, a photo was provided for review.The investigation of the reported event is currently underway.(expiry date: 08/2024).
 
Event or Problem Description
It was reported that during an angioplasty procedure, the pta balloon was allegedly ruptured.It was further reported that the balloon distal part was allegedly detached from the proximal ends of the balloon.The procedure was completed using another device.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one photo was provided and reviewed.The photo shows a detached portion of the catheter with the inner guide-wire lumen and balloon connected to the distal tip.One pta dilatation catheter was received for evaluation in two segments.During visual evaluation, segment 1 was noted to consist of the inner guide-wire lumen and the balloon.The balloon was still connected at the distal tip and inverted.Segment 2 consisted of a partial section of the catheter.No functional testing was performed due to the condition of the returned sample.Therefore, the investigation is inconclusive for the reported balloon rupture as no evidence of balloon rupture could be determined.The investigation is confirmed for the reported detachment as the catheter was noted to be detached from both the photo review and returned sample.A definitive root cause for the reported balloon rupture and detachment could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 08/2024).H11: h6 (method, result, conclusion).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during an angioplasty procedure, the pta balloon was allegedly ruptured.It was further reported that the balloon distal part was allegedly detached from the proximal ends of the balloon.The procedure was completed using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATLAS GOLD
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
FUTUREMATRIX INTERVENTIONAL
1605 enterprise street
athens 75751
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key14629912
Report Number2020394-2022-00449
Device Sequence Number12306233
Product Code DQY
UDI-Device Identifier00801741060823
UDI-Public(01)00801741060823
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K181323
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberATG120164
Device Catalogue NumberATG120164
Device Lot Number93VF0128
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/18/2022
Supplement Date Received by Manufacturer08/01/2022
Initial Report FDA Received Date06/08/2022
Supplement Report FDA Received Date08/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-