SYNTHES GMBH DURA GUARD-MEDIUM FOR PEN DRIVE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
|
Back to Search Results |
|
| Catalog Number |
05.001.052 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/31/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by switzerland that during service and evaluation, it was determined that the craniotome neuro-tip was fractured in 2 pieces.It was noted in the service order that during test drive, it was discovered that the device was broken.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Device evaluation: this device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, an evaluation was performed and it was determined that the reported condition of craniotome neuro-tip being fractured identified during service and evaluation is confirmed.The assignable root cause was determined to be traced to user, which is user error.Udi: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|