BIOMET UK LTD. BIOLOX DELTA HEAD, 36MM +3MM NECK, TYPE 1 TAPER; PROSTHESIS, HIP, SEMI-CONSTRAINED Back to Search Results
Model Number 650-0662
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Pain (1994); Joint Dislocation (2374)
Date of Event 04/27/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Concomitant medical device: acetabular shell g7 3 hole, pps; catalog#010000664; lot#7167132, trilogy self-tapping bone screw; catalog#63759079; lot#63759079, acetabular g7 liner, longevity; catalog#20103606 ; lot#65255998, femoral stem echo bi-metric full proximal profile, lateralized; catalog#192111; lot#501010.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event or Problem Description
It was reported that the patient was bending and twisting and felt a pop and increased pain on left hip approximately 1,5 month after initial left hip procedure.Anterior dislocation was confirmed; patient underwent a reduction under anesthesia.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Review of the device history records identified no deviations or anomalies during manufacturing that could be related to the reported event.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: an initial left tha was performed, later the patient was bending and twisting and felt a sudden pop with pain.Attempts at closed reductions were unsuccessful in the er, so the hip was reduced under anesthesia.No fractures noted.It was reported the patient was bending and twisting and felt a sudden pop.However, a root cause cannot be identified.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOLOX DELTA HEAD, 36MM +3MM NECK, TYPE 1 TAPER
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
Manufacturer (Section G)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK   CF31 3XA
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key14631121
Report Number3002806535-2022-00268
Device Sequence Number17050453
Product Code LZO
UDI-Device Identifier00887868271182
UDI-Public(01)00887868271182(17)311028(10)3090731
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K192683
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number650-0662
Device Catalogue Number650-0662
Device Lot Number3090731
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/09/2022
Supplement Date Received by Manufacturer06/16/2022
Initial Report FDA Received Date06/08/2022
Supplement Report FDA Received Date07/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/28/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10 NARRATIVE.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age69 YR
Patient SexMale
Patient Weight86 KG
-
-