Model Number 0010-0210R
Medical Device Problem Code Device Displays Incorrect Message (2591)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/19/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the console displayed error codes 630, 302.13, 102.3, 202.11 and 710.A problem with the chiller was suspected.It was indicated that the procedure could not be completed due to this event.There were no clinical consequences to the patient.This event is being reported for aborted/cancelled procedure with a patient whose sedation status was unknown.
 
Additional Manufacturer Narrative
The console was not returned for analysis; however, this console was serviced at the customer's facility by a field service engineer (fse).The fse replaced the flow switch, and the reported errors were cleared.The laser was power tested at 20, 30 and 40 watts coagulation and 80, 120 and 180 watts vaporization.The aim beam and interlock functionality were verified and the laser performed within specifications.The review completed on the device history record (dhr) for this console confirmed that it met specification prior to release, and the most recent console inspection found it to be fully functional with no issues.Based on the information available, it was concluded that the flow switch was the most probable cause of the reported event.
 
Event or Problem Description
It was reported that the console displayed error codes 630, 302.13, 102.3, 202.11 and 710.A problem with the chiller was suspected.It was indicated that the procedure could not be completed due to this event.There were no clinical consequences to the patient.This event is being reported for aborted/cancelled procedure with a patient whose sedation status was unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GREENLIGHT XPS LASER SYSTEM
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
alyson harris
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key14631195
Report Number2124215-2022-20191
Device Sequence Number14362268
Product Code GEX
UDI-Device Identifier00878953001906
UDI-Public00878953001906
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K092735
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 07/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0010-0210R
Device Catalogue Number0010-0210R
Was Device Available for Evaluation? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/19/2022
Supplement Date Received by Manufacturer06/22/2022
Initial Report FDA Received Date06/08/2022
Supplement Report FDA Received Date07/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/11/2013
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexMale
-
-