It was reported that a progav (part # fv413t) was implanted during a procedure performed on an unknown date.According to the complainant, the shunt system was believed to be operating in overdrainage and had adjustment difficulties.The patient underwent a revision procedure on an unknown date.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: 11 years.Height: 140 centimeters (cm).Weight: 36,9 kilograms (kg).Gender: male.Same event as medwatch-no.3004721439-2022-00147.
|
|
Investigation: visual inspection: a deformation of the outer housing of the progav valve was observed through the visual inspection.The measurement of the plane parallelism could confirm that with a value of -0,033mm the housing membrane is clearly outside tolerance (0 ± 0.02 mm).Permeability test: a permeability test has shown that the valve is permeable.Adjustment test: the progav was tested and is adjustable to all specified pressures.Klick force and brake function test: the brake functionality test has shown that the brake function is fully operational and the braking force is within the given tolerances.Computer controlled test: to investigate the claim of over-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valve is tested in both the horizontal as well as the vertical positions.The results show that the valve operates within the accepted tolerance in the horizontal position.Internal inspection: after dismantling of the valve, deposits were found in progav.Result: for the sake of thoroughness and transparency, we would like to point out that an adjustment test and permeability test was already carried out after the revision at the hospital.Based on our investigation results, we can determine deposits.The deposits had no impact to the technical properties at the time of the investigation.The cause of the aforementioned functional impairment is not known to us at the time of the examination.Proteins in the cerebrospinal fluid can influence the function temporarily and are known side effects in hydrocephalus therapy.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
|