(b)(4).Common device name: device product code - phx.Concomitant medical products: item# 115310; lot# j7207291, item# 405889; lot# 810610, item# 405880; lot# 531830, item# 110027734; lot# 937180, item# 110031378; lot# 937190, item# 180555; lot# 030310, item# 180555; lot# 819850, item# 180554; lot# 161130, item# 180554; lot# 095380, item# 115399; lot# 937950.Initial reporter name and address: full establishment name - (b)(6) health campus.Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the product location is unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected: b4; b5; g3; g6; h1; h2; h3; h6.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Medical records were not provided.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.The root cause of the reported issue is attributed to patient noncompliance.According to doctor, patient was mowing his lawn and had a hard jerking motion that caused the dislocation after the first revision.Post operatively the shoulder should be protected, in a sling until cleared by the surgeon.Using the arm to mow the lawn within 2 weeks postoperatively is considered noncompliance with surgeons instructions.If any further information which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|