Model Number 710302
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/09/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Following the reported event, the device was sent to the manufacturer for evaluations and repairs.The coefficient endoscopic insufflator® system operator's manual (732223) gives the user the following information if the unit does not deliver co2.Possible cause: "coefficient endoscopic insufflator flow rate incorrect.To check the coefficient endoscopic insufflator's ability to deliver co2 to the patient, follow these verification steps: after powering-up and making sure that your coefficient endoscopic insufflator is connected to the co2 supply tank, depress the volume reset button, zero will appear in the volume liters display.Set the mode selector switch to free flow.Disconnect the tubing set if connected.Press the flow stop/run button to start.Using a wrist watch with a second hand or a stop watch, allow the unit to deliver 3.0 liters of co2.The unit should reach this volume between 50 and 70 seconds.During this time interval, place your finger near the gas output.You should feel co2 exiting from the coefficient endoscopic insufflator".Steris is pending the report of the evaluations.A follow-up report will be submitted when additional information becomes available.No additional issues have been reported.
 
Event or Problem Description
The user facility reported that during a patient procedure the flow rate on their coefficient endoscopic insufflator system was low.A procedure delay was reported due to the low flow rate.The procedure was completed successfully and no injuries were reported.
 
Additional Manufacturer Narrative
In lieu of having the device return to be repaired at steris, the customer spoke with a steris product specialist and were retrained on the instructions concerning flow rate.The customer confirmed that the device was operating to specifications and returned to service.No additional issues have been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CO2EFFICIENT ENDOSCOPIC INSUFFLATOR SYSTEM
Common Device Name
CO2EFFICIENT ENDOSCOPIC INSUFFLATOR SYSTEM
Manufacturer (Section D)
UNITED STATES ENDOSCOPY GROUP, INC.
5976 heisley road
mentor OH 44060
Manufacturer (Section G)
UNITED STATES ENDOSCOPY GROUP, INC.
5976 heisley road
mentor OH 44060
Manufacturer Contact
coletta cohara
5976 heisley road
mentor, OH 44060
4403586251
MDR Report Key14631778
Report Number1528319-2022-00030
Device Sequence Number14431781
Product Code FCX
UDI-Device Identifier10816765013097
UDI-Public10816765013097
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number710302
Device Catalogue Number710302
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/09/2022
Supplement Date Received by Manufacturer05/09/2022
Initial Report FDA Received Date06/08/2022
Supplement Report FDA Received Date08/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/01/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-