ETHICON INC. SURGICEL POWDER USA - ENDOSCOPIC APPLICATOR; AGENT, ABSORBABLE HEMOSTATIC, NON-COLLAGEN BASED
|
Back to Search Results |
|
| Model Number |
3123SPEA |
| Medical Device Problem Code |
Delivered as Unsterile Product (1421)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Event related to mw # 2210968-2022-04375, mw # 2210968-2022-04376, mw # 2210968-2022-04377, mw # 2210968-2022-04379.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that before a patient underwent an unknown procedure on an unknown date, upon arrival the absorbable hemostat applicator was noticed to be unsealed.No adverse patient consequences were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary = > device history lot = > "a manufacturing record evaluation was performed for the finished device batch sabexl , 3123spea and no non-conformances were identified.Mfg.Date - 1/28/2022 exp.Date - 12/31/2026" product complaint # (b)(4).Corrected information: h3, h6.Component code corrected h3 investigation summary = > the product was returned for evaluation.Visual inspection was conducted on the returned samples.Visual analysis of the returned sample revealed that one bottom overwrap and one surgicel endoscopic applicator of product code 3123spea were returned.The package overwrap was received opened and the top overwrap was not returned for analysis.The seal area of the bottom overwrap was inspected and the seal print evidence could be observed within the perimeter of the overwrap, suggesting that the package was properly sealed.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.As the complete packaging material was not returned for analysis, it is not possible to determine what caused the reported event.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).Additional information: d9.Date device returned to manufacturer h3 investigation summary one bottom over wrap and a one surgicel endoscopic applicator of product code 3123spea was returned to ethicon for analysis.In order to evaluate the conditions of the returned sample, the sample was received completely open.For this reason, it was not possible determine if, the applicators were unsealed.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.No conclusion could be reached as to what caused the reported complaint.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected information: d9.Device available for evaluation, h3, h6.
|
| |
|
Search Alerts/Recalls
|
|
|