Model Number FX576T
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 05/06/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation: visual inspection: bloody residues were found on the incoming product.Permeability test: a permeability test has shown that both valves are permeable.Computer controlled test: to investigate the claim of under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valves are tested in both the horizontal as well as the vertical positions.The results show that the valves operate within the accepted tolerance in both positions.Adjustment test: the progav 2.0 was tested and is adjustable to all specified pressures.Braking force and brake function test: the brake functionality test has shown that the brake function is fully operational and the braking force is within the given tolerances.Internal inspection: after dismantling of the valves, deposits were found in both valves.Results: it should be noted that the product was received dry (i.E.Not submersed in liquid as recommended).The investigation of dry items is not significant due to the affect dry deposits of liquor and blood can have on product performance.In spite of this, we have investigated the valve to the best of our abilities.For the sake of thoroughness and transparency, we would like to point out that a permeability test was already carried out after the revision at the hospital.Based on our examination results, we are able to determine deposits.The deposits had no influence on the technical characteristics at the time of the examination.We are unable to determine how the aforementioned functional impairment occurred at the time of the investigation.Existing deposits in the csf can temporarily impair the function of the valves and is described as concomitant symptoms in hydrocephalus therapy.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke (b)(4).No further regulatory actions are required from our point of view.
 
Event or Problem Description
It was reported that a progav 2.0 shunt system (part # fx576t) was implanted during a procedure performed on (b)(6) 2021.According to the complainant, the shunt system was believed to be operated in underdrainage.The patient underwent a revision procedure on may 6, 2022.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: 16 years, height: 178 centimeters (cm), weight: 54 kilograms (kg), gender: male.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGAV 2.0 SYS SA2.0 20 & SPRUNG RESERV.
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key14632422
Report Number3004721439-2022-00152
Device Sequence Number17621681
Product Code JXG
UDI-Device Identifier04041906502522
UDI-Public4041906502522
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K190174
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/02/2026
Device Model NumberFX576T
Device Catalogue NumberFX576T
Device Lot Number20053722
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/12/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/12/2022
Initial Report FDA Received Date06/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/23/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age16 YR
Patient SexMale
Patient Weight54 KG
-
-