Model Number LOGIC TIBIA IMPLANT PSC INSERT, SZ 1, 13MM
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 05/17/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Event or Problem Description
As reported, approximately two years post initial left tka, the 66 y/o female patient had a revision.Patient was complaining since 6 months of original tka of pain, swelling and discomfort.Patient had little inflammatory syndrome but presence of granulum.The pe is heavily worn behind the posterior cam and in the angle of the external femoral condyle.Only patient's insert has been replaced.X-rays and photos received.The device will be return.
 
Additional Manufacturer Narrative
After further review of additional information received the following sections d9, g3, g6, h2, h3, h6 and h9 have been updated accordingly.(h3) the revision reported was likely the result of joint laxity and/or instability which led to the femoral cam contacting the posterior of the tibial insert spine and subsequent wear.Additionally, it is likely that a contributing factor to the wear seen on the returned tibial insert may have been the result of being packaged in a non-conforming vacuum bag for more than five years.The most probable root cause associated with the reported event of "prosthesis wear¿ is associated with material damage to a surface, usually involving progressive loss or displacement of material, due to relative motion between that surface and a contacting substance or substances.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC TIBIA IMPLANT PSC INSERT, SZ 1, 13MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED,
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
kate jacobson
2320 nw 66 court
gainesville, FL 32653
3523771140
MDR Report Key14632498
Report Number1038671-2022-00651
Device Sequence Number17621972
Product Code JWH
UDI-Device Identifier10885862173416
UDI-Public10885862173416
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K110547
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/13/2023
Device Model NumberLOGIC TIBIA IMPLANT PSC INSERT, SZ 1, 13MM
Device Catalogue Number02-012-44-1013
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/19/2022
Supplement Date Received by Manufacturer09/20/2022
Initial Report FDA Received Date06/08/2022
Supplement Report FDA Received Date10/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/14/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0021-2022
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
Patient SexFemale
-
-