| Model Number |
IPN919609 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device has not been returned for investigation.
|
| |
|
Event or Problem Description
|
|
This epidural kit was opened and contained an open, empty glass ampule that should have contained saline.The solution appeared to have dried and nothing was wet in the kit.This is a sharps issue due to the opened glass ampules and a quality issue.
|
| |
|
Event or Problem Description
|
|
This epidural kit was opened and contained an open, empty glass ampule that should have contained saline.The solution appeared to have dried and nothing was wet in the kit.This is a sharps issue due to the opened glass ampules and a quality issue.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).A device history record review was performed on the kit and the ampules with no relevant findings.A corrective action is not required at this time as a potential root cause could not be determined based upon the information provided and without a sample.Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed on the kit and ampules with no evidence to suggest a manufacturing related cause.Therefore, the potential cause of this complaint could not be determined based upon the information provided and without a sample.No further action is required at this time.
|
| |
|
Search Alerts/Recalls
|