• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 NEUTRAL ARCOMXL LNR 36MM E; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 NEUTRAL ARCOMXL LNR 36MM E; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/17/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported the liner would not seat.After three separate unsuccessful attempts, a new liner was dispensed and was implanted with zero issues.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
(b)(4).Concomitant medical products: 010000663- g7 pps ltd acet shell 52e- 7147518.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
No additional information on the reported event.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: d9; g3; h2; h3; h6 a g7 neutral arcomxl lnr 36mm e, part # 010000740 from lot 6629081, was returned and evaluated against the complaint.Visual inspection found the outer radius to be in decent overall shape with no scuffing or indentations.The barb is scraped in 2 locations.One of the locations has been scraped such that a poly strand protrudes from the barb.Gouging is present on the side wall around 1 of the scallops.A review of the device history records identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 NEUTRAL ARCOMXL LNR 36MM E
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key14633174
Report Number0001825034-2022-01370
Device Sequence Number12246616
Product Code PBI
UDI-Device Identifier00880304524972
UDI-Public(01)00880304524972(17)240930(10)6629081
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K121874
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2024
Device Model NumberN/A
Device Catalogue Number010000740
Device Lot Number6629081
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/17/2022
Supplement Date Received by Manufacturer06/16/2022
Initial Report FDA Received Date06/08/2022
Supplement Report FDA Received Date06/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/30/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
NI
Patient Sequence Number1
Patient SexFemale
-
-