• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH BEND-TEMPL F/MATRIXMAND-PREF-RECO-PL MED; TEMPLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH BEND-TEMPL F/MATRIXMAND-PREF-RECO-PL MED; TEMPLATE Back to Search Results
Catalog Number 03.503.911
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional narrative: date of event: unknown.Initial reporter is j&j company representative.Investigation summary: the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that bend-templ f/matrixmand-pref-reco-pl med was observed with signs of usage, the body section bent and the ramus broken.The broken ramus part was received in the evidence provided.A dimensional inspection for the bend-templ f/matrixmand-pref-reco-pl was unable to be performed due to post manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the bend-templ f/matrixmand-pref-reco-pl med would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed: no issues dimensional inspection: n/a.Device history: part # 03.503.911, lot # 8145510, manufacturing site: (b)(4), release to warehouse date: 18 feb2015, supplier: diener ag precision machining.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in columbia as follows: it was reported there were 2 matrixmandible templates that would not function.Specific malfunction was not reported.It was unknown when the devices malfunctioned.There is no known patient or surgical involvement.This is report 2 of 2 for (b)(4).This report is for a matrixmandible bndng template.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BEND-TEMPL F/MATRIXMAND-PREF-RECO-PL MED
Common Device Name
TEMPLATE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES SELZACH
bohackerweg 5
selzach 2545
SZ   2545
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key14633397
Report Number8030965-2022-03808
Device Sequence Number20313060
Product Code HWT
UDI-Device Identifier07611819363250
UDI-Public(01)07611819363250
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.503.911
Device Lot Number8145510
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/03/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/18/2022
Initial Report FDA Received Date06/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/18/2015
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-