(b)(4).The customer returned one loose clip, one empty cartridge, one lidstock, and two unopened representative samples of 001200 horizon ti small 6/cart 180/box for investigation.The returned samples were visually examined with and without magnification.Visual examination of the returned samples revealed that the clips appear typical.No defects or anomalies were observed.Functional inspection was performed on the returned samples.A lab inventory clip applier was used.All clips were able to properly load into the applier.The clips were able to close properly when applied to over-stressed surgical tubing and did not reopen.No functional issues were found with the returned clips.It is possible that the reported issue was caused by using damaged/misaligned appliers, but this could not be confirmed since the applier was not returned.The ifu for this product, l02428, was reviewed as a part of this complaint investigation.The ifu states, "always check the alignment of the applier jaws before use.When closed, jaw tips should be directly aligned and not offset.Alignment of the jaw is critical for safe application of the clip.If this is not done, patient injury may occur.Proper maintenance, care and cleaning are necessary to ensure proper functionality." a corrective action is not required at this time as there were no functional issues found with the samples that were returned.The reported complaint of "ligation failure-clip reopening" could not be confirmed based on the returned samples.One loose clip, one empty cartridge, one lidstock, and two unopened representative samples were returned.Upon functional inspection, no problems were found as clips could be loaded into the jaws of a lab inventory applier and close when applied to over-stressed surgical tubing without reopening.No functional issues were found with the returned clips.It is possible that the reported issue of "ligation failure-clip reopening" was caused by using damaged/misaligned appliers, but this could not be confirmed since the applier was not returned.Since no functional issues were found with the returned clips and the applier was not returned, the reported issue could not be confirmed.
|