MAUDE Adverse Event Report: MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number 3507555MC
Medical Device Problem Codes Gel Leak (1267); Material Rupture (1546); Migration (4003)
Health Effect - Clinical Code Deformity/ Disfigurement (2360)
Date of Event 04/05/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a (b)(6) caucasian female patient underwent a breast surgery with implantation of a 555cc mentor memorygel breast implant prosthesis.Post-operatively, the patient was diagnosed with a ruptured right breast implant using ultrasound imaging.As a result, the patient underwent removal and replacement with 675cc mentor memorygel breast implants on (b)(6) 2022.
 
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: rupture.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On june 17, 2022, the mentor analysis lab received the suspect medical device for evaluation.On june 23, 2022, mentor completed an evaluation on the returned device.Mentor conducted a visual inspection and leak testing of the device.Visual analysis of the returned sample revealed that no damage or anomalies were observed on the breast implant.Leak testing was performed, in accordance with mentor procedures, and no leak sites were detected during the analysis.Although no conclusion could be reached on the cause of the reported event, the instructions for use contain the following caution: rupture status identified by mri evaluation includes both suspected ruptures, which are those ruptures identified by mri but not confirmed by explantation and examination of the device, and confirmed ruptures, which are those ruptures that are confirmed by evaluation of the explanted devices.The event described could not be confirmed as the breast implant was returned without detectable damage.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On november 24, 2022, mentor received additional information.Mammogram imaging results indicated that a medial palpable silicone lump of the right breast was present.However, during removal surgery, the right implant was intact and a such, rupture is no longer applicable to this file.Since a lump was present without rupture, gel bleed is appropriate.Additionally, malposition of the right implant was indicated.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On december 15, 2022, mentor reevaluated the device per the updated adverse event.The following was added to the device evaluation.Leak testing was performed, in accordance with mentor procedures, and no leak sites or gel bleed were detected during the analysis.Implant displacement/migration may occur from improper implant sizing and/or placement, i.E., when the implant is too large or the pocket too small or when there has been inadequate preoperative assessment of stresses causing movement of the prosthesis.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
michelleann garcia
3041 skyway circle north
irving, TX 75038
646591-7981
MDR Report Key14633743
Report Number1645337-2022-06675
Device Sequence Number17637471
Product Code FTR
UDI-Device Identifier00081317004257
UDI-Public00081317004257
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/12/2020
Device Model Number3507555MC
Device Catalogue Number3507555MC
Device Lot Number6912850
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/18/2022
Supplement Date Received by Manufacturer06/17/2022
11/24/2022
12/15/2022
Initial Report FDA Received Date06/08/2022
Supplement Report FDA Received Date07/05/2022
12/14/2022
12/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/07/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age48 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-