• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDINAL HEALTH BALLOON SLG 20FR X 2.0CM; TUBES, GASTROINTESTINAL (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH BALLOON SLG 20FR X 2.0CM; TUBES, GASTROINTESTINAL (AND ACCESSORIES) Back to Search Results
Model Number 720200
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/02/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event or Problem Description
Customer reports: there was a balloon disruption after 16 days of use.
 
Additional Manufacturer Narrative
The device history record (dhr) was reviewed, and no abnormal process conditions were present during the manufacturing of the product that could have led to the reported condition from the customer complaint.The dhr review showed that all acceptance criteria inspections per established sampling levels were within acceptable limits during the production process.A sample analysis was not possible because there were no photographs or physical samples provided for evaluation.Without a sample analysis, we are unable to confirm the reported condition.The root cause and action plan will be documented through a formal corrective/preventative action (capa) which has been initiated with the supplier to address the reported condition to mitigate any further occurrences.
 
Additional Manufacturer Narrative
The device history record (dhr) review showed that all acceptance criteria inspections per established sampling levels were within acceptable limits during the production process.There were no deviations and no non-conformances initiated with this lot code.One decontaminated device within the original packaging and lot 192915064 was received for evaluation.A visual inspection was performed according to procedure.Additionally, a functional test was conducted by inflating the balloon with water and a leak was detected in the tubing of the balloon, but there was no rupture found in the balloon.The root cause and action plan will be documented through a formal corrective/preventative action (capa) which has been initiated to address the reported condition to mitigate any further occurrences.This action is currently ongoing.This complaint will be closed with no further action and will be used for tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BALLOON SLG 20FR X 2.0CM
Common Device Name
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
CARDINAL HEALTH
777 west street
mansfield MA 02048
Manufacturer (Section G)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX   22500
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key14633865
Report Number9612030-2022-03271
Device Sequence Number14932151
Product Code KNT
UDI-Device Identifier10884521011274
UDI-Public10884521011274
Combination Product (Y/N)N
Initial Reporter CountryBR
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number720200
Device Catalogue Number720200
Device Lot Number1929415064
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/23/2022
Supplement Date Received by Manufacturer05/23/2022
05/23/2022
Initial Report FDA Received Date06/08/2022
Supplement Report FDA Received Date06/10/2022
09/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-