• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD HEMI HEAD 42MM; PROSTHESIS, HIP, HEMI-, FEMORAL, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD HEMI HEAD 42MM; PROSTHESIS, HIP, HEMI-, FEMORAL, METAL Back to Search Results
Catalog Number 74122542
Medical Device Problem Code Biocompatibility (2886)
Health Effect - Clinical Codes Pain (1994); Metal Related Pathology (4530)
Date of Event 01/20/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, after left bhr-tha surgery was performed on (b)(6) 2009, the patient experienced pain at the hip joint and elevated cobalt and chromium levels.This adverse event was treated with revision surgery on (b)(6) 2021.During this procedure, the bhr modular head was removed from the system and replaced with a smith and nephew dual mobility liner and a 28mm +4 oxinium head.Intraoperatively, the acetabular component was noted to be in good position and without any evidence of loosening.The trunnion presented a small amount of corrosion in the mid distal region but was overall in good shape.The patient was transferred to the post anesthesia care unit without any complications.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
H3, h6: it was reported that left hip revision surgery was performed.As of today, the implanted devices, all of which were used in treatment have not been returned for evaluation.A review of the historical complaints data for the acetabular cup, hemi head and sleeve was performed using batch numbers to evaluate patterns of repeated failures or defects over the lifetime of the product and using part numbers and the reported failure modes to evaluate patterns of repeated failures or defects in a timeframe prior 12 months as of the complaint aware date.No other similar complaints were identified to involve this batch of acetabular cup and hemi head.Other similar complaints were identified to involve this batch of sleeve.Other similar complaints have been identified for the part number of acetabular cup and hemi head and the reported failure mode, and this failure will continue to be monitored.No other similar complaints were identified for the part number and the reported/related failure mode for the sleeve.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.All released devices involved met manufacturing specifications upon release for distribution.¿ review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.No further actions are required at this time.A review of historic escalation actions related to the products and similar complaint events was performed.Following the review, prior applicable escalation actions were identified and confirmed to reduce associated risks as far as possible.No further escalation actions required.The available medical documents were reviewed.Although it was reported that the revision surgery was performed secondary to pain and elevated cobalt and chromium, neither the levels nor the lab reports were provided for the review.It should be noted the surgical technique for bhr (01/07 4567-0103) indicates, the acetabular component is to be fully impacted with 15-20° of anteversion and 40-45° inclination angle.It is unknown if the increased anteversion of the acetabular component led to accelerated wear and the reported pain, elevated cobalt and chromium, and small amount of corrosion on the trunnion.It cannot be concluded that the reported clinical reactions were associated with a mal performance of the implant.The patient impact beyond the pain, revision and expected convalescence period cannot be determined.Based on the information provided we cannot confirm or further investigate the reported complaint, our investigation remains inconclusive, and a definitive root cause cannot be determined.Specific factors known to contribute to the alleged fault are excessive physical activity levels, excessive patient weight, trauma to the joint replacement.Should the devices or additional information be received, the complaint will be reopened.Based on the investigation, the need for corrective and preventative actions is not indicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMI HEAD 42MM
Common Device Name
PROSTHESIS, HIP, HEMI-, FEMORAL, METAL
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK  CV31 3HL
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK   CV31 3HL
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14634321
Report Number3005975929-2022-00387
Device Sequence Number5937144
Product Code KWL
UDI-Device Identifier00885556070710
UDI-Public885556070710
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K062408
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/30/2013
Device Catalogue Number74122542
Device Lot Number08HW18320
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/19/2022
Supplement Date Received by Manufacturer08/23/2022
Initial Report FDA Received Date06/08/2022
Supplement Report FDA Received Date08/23/2022
Date Device Manufactured08/31/2008
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ANTHOLOGY HO POR PL HA SZ 5, LOT#:09BM17828A; MODULAR SLEEVE +4MM 12/14, LOT#:08JW19038
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age68 YR
Patient SexFemale
-
-