Model Number 364489
Medical Device Problem Codes Difficult to Remove (1528); Failure to Advance (2524); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/10/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Combination product: yes.
 
Event or Problem Description
An orsiro drug-eluting stent system was selected for treatment of a moderately calcified lesion (70 percent stenosis degree) in the mid cx.After pre-dilatation it was not possible to advance the stent.Another pre-dilatation was performed, and the device was re-inserted but a deformation was noticed and the whole system was removed with difficulties from the patients body.
 
Additional Manufacturer Narrative
Combination product: yes.The returned instrument was subjected to a detailed technical analysis and the corresponding product release documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.The delivery system and the stent were returned separately.The stent was located inside a closed haemostatic valve.The stent was removed from the haemostatic valve and found severely deformed at its proximal end.The balloon is well folded and shows no signs of inflation.The proximal balloon shoulder is compressed.Microscopic inspection showed stent imprints on the exposed balloon surface, indicating that the stent was initially crimped centered on the balloon.Inspection of the hemostatic valve revealed no damage or irregularity.Review of the product release documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.As a part of final inspection every stent system undergoes visual inspection to ensure correct embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent and the stent retention force is tested in a defined amount of samples.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause was determined.The root cause for the complaint event is most likely related to the handling, i.E.Removal of the stent system through a not fully opened hemostatic valve which is warned against in the ifu.The ifu further advises the user to not re-insert the stent system as the stent and/or the delivery system may have been damaged during the initial attempt to cross the lesion or during withdrawal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORSIRO 2.75/18
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key14634564
Report Number1028232-2022-02969
Device Sequence Number5918337
Product Code NIQ
UDI-Device Identifier07640130439047
UDI-Public7640130439047
Combination Product (Y/N)Y
Initial Reporter CountryCH
PMA/510(K) Number
P170030
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/11/2022
Device Model Number364489
Device Catalogue NumberSEE MODEL NO.
Device Lot Number04201846
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/08/2022
Supplement Date Received by Manufacturer07/06/2022
Initial Report FDA Received Date06/08/2022
Supplement Report FDA Received Date07/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/11/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-