MAUDE Adverse Event Report: LAKE REGION MEDICAL ZIPWIRE HYDROPHILIC GUIDEWIRE; WIRE, GUIDE, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LAKE REGION MEDICAL ZIPWIRE HYDROPHILIC GUIDEWIRE; WIRE, GUIDE, CATHETER Back to Search Results
Model Number M00146151B1
Medical Device Problem Code Device Damaged by Another Device (2915)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/16/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
It was reported that the device will not return; therefore, no physical analysis of the device can be performed.A review of the device history records of the specimen device does not present any indication of manufacturing defect or anomaly that could have impacted on the event as reported.During manufacturing the production operators are instructed to 100% visually and tactilely inspect for any obvious defect, which includes a tactile examination of the entire length of each wire.In addition, during packaging of this product the operators are instructed to 100% visually inspect for any obvious defect prior to shipment.The history records indicate this product was final inspection tested at lake region medical and was determined to be acceptable.As indicated in the device instructions for use, warnings: do not manipulate or withdraw the zipwire hydrophilic guide wire through a metal entry needle or metal dilator.Manipulation and/or withdrawal through a metal entry needle or metal dilator may result in destruction and or separation of the outer polyurethane coating, requiring retrieval.A plastic entry needle is recommended when using this wire for initial placement.To prevent possible tissue damage, care should be taken when manipulating a device over a zipwire hydrophilic guide wire during the device's placement or withdrawal.If resistance is felt during device placement, discontinue the procedure and determine the cause of resistance before proceeding.If the cause of resistance cannot be determined, remove the zipwire hydrophilic guide wire and device as a unit to prevent possible damage and/or complications.If any resistance is felt or if the tip's behavior and/or location seems improper, stop manipulating the zipwire¿ hydrophilic guide wire and/or catheter and determine the cause under fluoroscopy.Failure to exercise proper caution may result in bending, kinking, separation of the guide wire's tip, damage to the catheter or damage to the vessel." at this time it is not possible to assign a definitive root cause for the event as reported.If there is any further relevant information provided, a follow up medwatch report will be submitted.
 
Event or Problem Description
Event description: it was reported that on (b)(6), when the guide wire was used for angiography, it could not be delivered into the angiography catheter normally.After examination, the physician found that the tip of the guide wire was bifurcated and could not be used normally.The physician replaced a new guide wire as soon as the physician found the problem.
 
Additional Manufacturer Narrative
It was reported that the device will not return; therefore, no physical analysis of the device can be performed.A review of the device history records of the specimen device does not present any indication of manufacturing defect or anomaly that could have impacted on the event as reported.During manufacturing the production operators are instructed to 100% visually and tactilely inspect for any obvious defect, which includes a tactile examination of the entire length of each wire.In addition, during packaging of this product the operators are instructed to 100% visually inspect for any obvious defect prior to shipment.The history records indicate this product was final inspection tested at lake region medical and was determined to be acceptable.According to the additional information provided 14 june 2022, the zipwire was advanced through a metal cannula.As indicated in the device instructions for use, warnings: ·do not manipulate or withdraw the zipwire hydrophilic guide wire through a metal entry needle or metal dilator.Manipulation and/or withdrawal through a metal entry needle or metal dilator may result in destruction and or separation of the outer polyurethane coating, requiring retrieval.A plastic entry needle is recommended when using this wire for initial placement.·to prevent possible tissue damage, care should be taken when manipulating a device over a zipwire hydrophilic guide wire during the device's placement or withdrawal.If resistance is felt during device placement, discontinue the procedure and determine the cause of resistance before proceeding.If the cause of resistance cannot be determined, remove the zipwire hydrophilic guide wire and device as a unit to prevent possible damage and/or complications.·if any resistance is felt or if the tip's behavior and/or location seems improper, stop manipulating the zipwire¿ hydrophilic guide wire and/or catheter and determine the cause under fluoroscopy.Failure to exercise proper caution may result in bending, kinking, separation of the guide wire's tip, damage to the catheter or damage to the vessel." if there is any further relevant information provided, a follow up medwatch report will be submitted.
 
Event or Problem Description
Event description: it was reported that: on may.16th, when the guide wire was used for angiography, it could not be delivered into the angiography catheter normally.After examination, the physician found that the tip of the guide wire was bifurcated and could not be used normally.The physician replaced a new guide wire as soon as the physician found the problem.Additional information received 14 june 2022: 1.Was it the first time the device was used? ans: no.2.Was there visible damage to the device and/or its packaging prior to use? ans: no.3.Was the guidewire hydrated or rinsed with saline solution prior to removal from the dispenser? ans: yes 4.Was there an issue with lubricity of the device? ans: not note.5.Any resistance felt during insertion? ans: no 6.Was the zipwire advanced through a metal cannula or needle? ans: a metal cannula 7.The narrative states, "the tip of the guide wire was bifurcated," is the metal core wire exposed? ans: no.8.Is there an image available? ans: no.9.Was there any patient injury reported? ans: no.10.Were there any patient complications during the procedure(s) or after as a result of the product issue? ans: no.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIPWIRE HYDROPHILIC GUIDEWIRE
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
LAKE REGION MEDICAL
340 lake hazeltine drive
chaska MN 55318
Manufacturer (Section G)
VENUSA DE MEXICO S. DE R.L. DE C.V.
calle hertz 1525
parque industrial a j bermudez
ciudad juarez, chihuahua 32470
MX   32470
Manufacturer Contact
victoria foss
340 lake hazeltine drive
chaska, MN 55318
9526418506
MDR Report Key14634636
Report Number9680001-2022-00104
Device Sequence Number17673679
Product Code DQX
UDI-Device Identifier08714729754695
UDI-Public08714729754695
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K000011
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/01/2024
Device Model NumberM00146151B1
Device Catalogue Number41496
Device Lot NumberJRZ6795140
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/25/2022
Supplement Date Received by Manufacturer05/25/2022
Initial Report FDA Received Date06/08/2022
Supplement Report FDA Received Date06/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/02/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age64 YR
Patient SexMale
-
-