| Model Number |
MAJ-1414 |
| Medical Device Problem Code |
Device Fell (4014)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/11/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The subject device has not yet been received for evaluation.The cause of the reported issue is unknown at this time.Supplemental report(s) will be submitted should any relevant new information is available.Investigation is ongoing.This report will be supplemented accordingly following investigation.
|
| |
|
Event or Problem Description
|
|
As reported, the stopper flew off when the patient coughed.The intended procedure was a diagnostic procedure.Procedure was completed with the same device.There was no patient harm or injury reported , no user injury reported due to the event.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and correction to d4.D4: corrected the model number.The lot number is unknown.Customer confirmed the device was discarded.The legal manufacturer performed internal testing by combining the company's scope (bf-uc260fw sn: (b)(6) and the company's forceps plug (maj-1414, lot: 24k).The testing was completed by forcibly applying pressure from the tip of the scope to check if the forceps plug came off but could not reproduce the removal of the forceps plug.Based on the investigation results, the legal manufacturer has determined there is no potential for this malfunction of the biopsy valve flying off in the room to cause or contribute to an adverse event.H3 other text: device was discarded.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on customer response and updates regarding event reported.Communication with the customer representative conveyed that the biopsy valve flew off in the room, not in patient and it was not inside the patient.No further information provided.Investigation is ongoing.This report will be supplemented accordingly following investigation.
|
| |
|
Search Alerts/Recalls
|