| Model Number |
466FXXXX |
| Medical Device Problem Codes |
Fracture (1260); Unintended Movement (3026)
|
| Health Effect - Clinical Codes |
Perforation (2001); Perforation of Vessels (2135)
|
| Date of Event |
08/16/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
As reported a patient underwent placement of an optease vena cava filter.The information provided indicated that the filter subsequently malfunctioned and caused filter fracture, organ perforation and tilt.The indication for the filter implant, procedural details and medical history of the patient have not been provided and there is currently no additional information available for review.The product remains implanted and unavailable for analysis, the sterile lot number has not been provided; therefore, no device analysis nor device history record review could be performed.The optease vena cava filter is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the vena cava for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pe where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pe where anticoagulant therapy has failed, or is contraindicated.The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart.Ivc filter tilt has been associated with operator technique and vessel anatomy, specifically asymmetry and tortuosity.Vessel perforation is a known adverse event associated with implanting vena cava filters and is listed as such in the instructions for use (ifu).The ifu also notes vessel damage such as intimal tears and perforation as procedural and long-term complications related to ivc filters.The ifu also states that filter fracture with perforation is a potential complication of vena cava filters.Anatomic locations that create concentrated stress points from filter deformation (for example, deployment at apex of scoliosis, overlapping of either of the renal ostia, or placement adjacent to a vertebral osteophyte) may contribute to fracture of a particular filter strut.The timing and mechanism of the events has not been reported at this time and a clinical conclusion could not be determined as to the cause of the events.Without images available for review the reported events could not be confirmed or further clarified.Given the limited information available for review at this time, there is nothing to suggest that the reported events are related to the design and manufacturing process of the device; therefore, no corrective action will be taken.Should additional information become available, the file will be updated accordingly.
|
| |
|
Event or Problem Description
|
|
As reported by the legal department, the patient underwent placement of the optease vena cava filter.The filter subsequently malfunctioned and caused injury and damages to the patient including, but not limited to, filter fracture, organ perforation and tilt.As a direct and proximate result, the patient suffered life-threatening injuries and damages and required extensive medical care and treatment.As a further proximate result, the patient has suffered and will continue to suffer significant medical expenses, pain and suffering, and other damages.
|
| |
|
Event or Problem Description
|
|
Additional information received per the medical records indicate that the patient has a history of morbid obesity with increased propensity for deep venous thrombosis and pulmonary emboli.The filter was implanted via the patient's right common femoral vein.A guide wire was passed to the 2nd and 3rd lumbar vertebrae.The introducing catheter was positioned at this level and an inferior venocavogram was done.The diameter of the inferior vena cava was found to be satisfactory and the filter was deployed under direct fluoroscopic guidance.There was no evidence of migration of the filter upon removal of the introducing catheter.Direct pressure was held over the puncture site in the right groin for 10 minutes.The patient tolerated the procedure well.Additional information received per the patient profile form (ppf) states that the patient experienced fracture in ivc, tilt, organ perforation and ivc perforation.The patient became aware of the reported events approximately thirteen years and four months after the index procedure.The patient also reported experiencing pain (stomach and leg), leg swelling (worse in left leg), anxiety and scared that filter will break off.
|
| |
|
Additional Manufacturer Narrative
|
|
As reported a patient underwent placement of an optease vena cava filter.The information provided indicated that the filter subsequently malfunctioned and caused filter fracture, organ perforation and tilt.The patient reported becoming aware of filter fracture tilt, inferior vena cava (ivc) and organ perforation, approximately thirteen years and four months post implant.The patient also reported stomach and leg pain, leg swelling, and anxiety related to the filter.According to the medical record the indication for the filter implant was a history of morbid obesity with increased propensity for deep venous thrombosis and pulmonary emboli.The filter was implanted via the right common femoral vein.And was deployed under direct fluoroscopic guidance.There was no evidence of migration of the filter upon removal of the introducing catheter.The patient tolerated the procedure well.The product remains implanted and unavailable for analysis.A review of the device history record revealed no anomalies during the manufacturing and inspection processes that can be associated with the reported complaint.The optease vena cava filter is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the vena cava for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pe where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pe where anticoagulant therapy has failed, or is contraindicated.The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart.Ivc filter tilt has been associated with operator technique and vessel anatomy, specifically asymmetry and tortuosity.Vessel perforation is a known adverse event associated with implanting vena cava filters and is listed as such in the instructions for use (ifu).The ifu also notes vessel damage such as intimal tears and perforation as procedural and long-term complications related to ivc filters.The ifu also states that filter fracture with perforation is a potential complication of vena cava filters.Anatomic locations that create concentrated stress points from filter deformation (for example, deployment at apex of scoliosis, overlapping of either of the renal ostia, or placement adjacent to a vertebral osteophyte) may contribute to fracture of a particular filter strut.The timing and mechanism of the events has not been reported at this time and a clinical conclusion could not be determined as to the cause of the events.Without images available for review the reported events could not be confirmed or further clarified.Pain, leg swelling, and anxiety do not represent a device malfunction and may be related to underlying patient specific issues and comorbidities.Given the limited information available for review at this time, there is nothing to suggest that the reported events are related to the design and manufacturing process of the device; therefore, no corrective action will be taken.Should additional information become available, the file will be updated accordingly.
|
| |
|