• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LUMINEX CORPORATION VERIGENE PROCESSOR SP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LUMINEX CORPORATION VERIGENE PROCESSOR SP Back to Search Results
Model Number 10-0000-07R
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 05/26/2022
Type of Reportable Event Malfunction
Event or Problem Description
The service visit at tallahassee memorial hospital was scheduled per (b)(4) within work order wo-00200320.During the site visit, the hybridization (hyb) heater required recalibration due to failing initial verification at setpoint 2.Field quality updated risk assessment question 2 to 'yes' since the setpoint temperature failed outside the accepted limits established by the verigene i assay temperature tolerance guard band risk report (03838).
 
Additional Manufacturer Narrative
In review of all associated customer-provided documentation and internally generated investigation notes against the criteria set forth in document 00043 (medical device reporting standards operating procedure), this complaint does meet the requirements for an mdr in the countries where this product is sold or distributed.Filing individual mdr out of abundance of caution due to the remote possibility of a false result contributing to a death or serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERIGENE PROCESSOR SP
Common Device Name
VERIGENE PROCESSOR SP
Manufacturer (Section D)
LUMINEX CORPORATION
12212 technology blvd
austin TX 78727
Manufacturer Contact
jennifer svoboda
12212 technology blvd
austin, TX 78727
MDR Report Key14637995
Report Number1650733-2022-00035
Device Sequence Number17383815
Product Code NSU
UDI-Device Identifier00840487101490
UDI-Public00840487101490
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number10-0000-07R
Device Catalogue Number10-0000-07R
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/26/2022
Initial Report FDA Received Date06/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/24/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-