TEPHA, INC. - 3005670760 GALAFLEX 3D; BREAST - ABSORBABLE MESH Back to Search Results
Model Number 100042-04
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code Impaired Healing (2378)
Date of Event 06/26/2018
Type of Reportable Event Serious Injury
Event or Problem Description
Per clinical trial cp-(b)(6) : on (b)(6) 2018, patient underwent bilateral mastopexy/reduction; superior pedicle using a full vertical and horizontal t incision.2 pieces of galashape (galaflex 3d) were used.Concomitant procedure was liposuction lateral breasts.On (b)(6) 2018 right breast delayed wound healing was reported.Patient treated with medication (silvadene 1% topical antibacterial cream and carrington wound gel).No debridement was necessary.Per notes: galashape (galaflex 3d) may have been exposed slightly at a point in recovery.Review of medical affairs identified that the delayed wound healing is likely related to the device and serious.
 
Additional Manufacturer Narrative
Based on the information provided, no conclusions can be made.As reported, patient experienced post-op complications of delayed wound healing.Medical affairs has assessed the patient¿s postoperative course as being serious and likely related to the study device.Review of the device history records confirmed the device was manufactured to specification.Note: this mdr report is related to a retrospective study conducted to proactively gather real-world evidence on the use and safety of the p4hb product family, including galaflex scaffold, galaflex 3d, and galaflex 3dr.All safety events from the study are adjudicated by the bd medical monitor for an independent assessment of severity and relatedness.In an abundance of caution, events for which relatedness cannot be ruled out are processed for mdr reporting.Remains implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALAFLEX 3D
Common Device Name
BREAST - ABSORBABLE MESH
Manufacturer (Section D)
TEPHA, INC. - 3005670760
99 hayden ave suite 360
lexington MA 02421
Manufacturer (Section G)
TEPHA, INC. - 3005670760
99 hayden ave suite 360
lexington MA 02421
Manufacturer Contact
andrew topoulos
100 crossings blvd.
warwick, RI 02886
8005566756
MDR Report Key14638174
Report Number1213643-2022-00408
Device Sequence Number5574930
Product Code OOD
UDI-Device Identifier00855920006171
UDI-Public(01)00855920006171
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K161092
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source Other,Study
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/17/2018
Device Model Number100042-04
Device Catalogue NumberSH3D04
Device Lot Number170582
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/31/2022
Initial Report FDA Received Date06/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/10/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age22 YR
Patient SexFemale
Patient Weight75 KG
-
-