Model Number 71421516
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Fall (1848); Failure of Implant (1924); Pain (1994); Joint Laxity (4526)
Date of Event 05/23/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, after a tka surgery had been performed on 2010, the patient fell on his tka.The patient presented with pain and instability in the clinic.A revision surgery was performed on the (b)(6) 2022 to treat this adverse event.During the surgery, it was revealed that the top post of the gii ps hi flex isrt sz 5-6 11 had fractured off and was lodged in soft tissue in the notch.All pieces were successfully removed and a new ps standard poly was implanted.The patient status is unknown.
 
Event or Problem Description
It was reported that, after a tka surgery had been performed on 2010, the patient fell on his tka.The patient presented with pain and instability in the clinic.A revision surgery was performed on the (b)(6) 2022 to treat this adverse event.During the surgery, it was revealed that the top post of the gii ps hi flex isrt sz 5-6 11 had fractured off and was lodged in soft tissue in the notch.All pieces were successfully removed and a new ps standard poly was implanted.The patient is recovering well.
 
Additional Manufacturer Narrative
The associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device was found to be fractured along the post, rendering the device inoperative.The device was also found to be discolored from exposure to bodily fluids.The fracture post was returned as well.A review of complaint history of the previous 12 months revealed similar events for the listed device, this failure mode will be monitored for future complaints for any necessary corrective actions.The clinical/medical evaluation concluded that the four undated, unlabeled photos, two of the inserts with the broken post, and two intra-operative photo of the open knee that revealed the broken insert, along with the chart sticks support the reported failure.The root cause of the reported failure could not be confirmed.The reported fall and the patient body habitus of 340 pounds are possible contributory factor; ¿the implant can break or become damaged because of strenuous activity or trauma and has a finite expected service life and may need to be replaced in the future.¿ it should be noted the implant was implanted 12 years prior to the revision and the patient is recovering well.The patient impact beyond the pain, revision, and expected transient post-op convalescence period cannot be determined.No further clinical assessment is warranted at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file and instructions for use document revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.As the device broke and it can no longer fit its purpose, the contribution of the device to the reported event could be corroborated.Some potential probable causes for this event could include but not limited to traumatic injury, abnormal loading of the limb, or implant failure.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII PS HI FLEX ISRT SZ 5-6 11
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14638969
Report Number1020279-2022-02814
Device Sequence Number12201703
Product Code JWH
UDI-Device Identifier03596010492333
UDI-Public03596010492333
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K032295
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71421516
Device Catalogue Number71421516
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/06/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/23/2022
Supplement Date Received by Manufacturer06/16/2022
07/15/2022
Initial Report FDA Received Date06/08/2022
Supplement Report FDA Received Date06/20/2022
07/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
Patient Weight154 KG
-
-