Model Number X070-0040
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/03/2022
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received notification of a system instrument malfunction on (b)(6) 2022.It was reported that the distal tip of a system screwdriver fractured while attempting to remove a system screw from an alternate manufacturer.There were no known patient complications associated with this complaint.The complaint screwdriver was received on 6/03/2022 at the manufacturer for complaint assessment.
 
Additional Manufacturer Narrative
A visual assessment of the returned system screwdriver showed an instrument with repeated use, as identified by worn laser markings and surface scratches.The distal tip of the driver was fractured and there were rotational wear marks around the remaining portion of the tip.A functionality assessment was not performed due to the damaged condition of the returned instrument, which was removed from distributable inventory.A dhr review was performed for the returned screwdriver complaint lot and there were no manufacturing anomalies identified.The device lot met all required specifications prior to being released to distributable inventory.This lot has been available for distribution since 2/18/2019.The complaint screwdriver is intended to be used with torque-limiting handles from the manufacturer to appropriately limit applied torque.It is unknown what required torque is appropriate for placing and removing the alternate manufacturer's system screws.The root cause of this complaint is excessive rotational force applied to the screwdriver to remove a system screw from an alternate manufacturer.There have not been any other complaints of similar nature in the past 12 months.The manufacturer will continue to monitor this instrument for complaints from the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORTEX PEDICLE SCREW SYSTEM
Common Device Name
PEDICLE SCREW SPINAL SYSTEM
Manufacturer (Section D)
XTANT MEDICAL HOLDINGS, INC.
664 cruiser lane
belgrade MT 59714
Manufacturer (Section G)
XTANT MEDICAL HOLDINGS, INC.
664 cruiser lane
belgrade MT 59714
Manufacturer Contact
rebecca lennemann
664 cruiser lane
belgrade, MT 59714
MDR Report Key14639066
Report Number3005031160-2022-00021
Device Sequence Number20097764
Product Code MNI
UDI-Device IdentifierM697X07000401
UDI-PublicM697X07000401
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K152132
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberX070-0040
Device Lot Number059814
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/03/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/19/2022
Initial Report FDA Received Date06/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/18/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-