BOSTON SCIENTIFIC CORPORATION ORIGEN X4 CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) Back to Search Results
Model Number G058
Medical Device Problem Codes Signal Artifact/Noise (1036); High impedance (1291); High Capture Threshold (3266); Noise, Audible (3273)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/19/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) system exhibited high out of range pacing impedance in the right ventricular (rv) and left ventricular (lv) leads.The device was also noted to exhibit beeping tones.The leads were measured with the pacing system analyzer (psa) and the lv lead measurements were noted to be in range.The rv lead measurements continued to be high and high threshold was also noted.It was mentioned that the rv lead could exhibit a lead conductor fracture or an insulation damage, and noise was also observed in this lead.Subsequently, the crt-d and rv lead were replaced and the new measurements after the products replacement were within adequate range.The lv lead remains in service.No additional adverse patient effects.The crt-d device is expected to be returned for analysis.Additional information indicates the crt-d device was replaced to be on the safe side due to the high out of range impedance measurements and the beeping tones observed were also caused by the high impedance measurements.In addition, the lv lead high out of range impedance is believed to be due to the stimulation being set against the rv lead.
 
Additional Manufacturer Narrative
The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.<
 
Event or Problem Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) system exhibited high out of range pacing impedance in the right ventricular (rv) and left ventricular (lv) leads.The device was also noted to exhibit beeping tones.The leads were measured with the pacing system analyzer (psa) and the lv lead measurements were noted to be in range.The rv lead measurements continued to be high and high threshold was also noted.It was mentioned that the rv lead could exhibit a lead conductor fracture or an insulation damage, and noise was also observed in this lead.Subsequently, the crt-d and rv lead were replaced and the new measurements after the products replacement were within adequate range.The lv lead remains in service.No additional adverse patient effects.The crt-d device is expected to be returned for analysis.Additional information indicates the crt-d device was replaced to be on the safe side due to the high out of range impedance measurements and the beeping tones observed were also caused by the high impedance measurements.In addition, the lv lead high out of range impedance is believed to be due to the stimulation being set against the rv lead.The device was returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORIGEN X4 CRT-D
Common Device Name
CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key14640120
Report Number2124215-2022-20291
Device Sequence Number5771312
Product Code NIK
UDI-Device Identifier00802526534515
UDI-Public00802526534515
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
P010012/S341
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date05/19/2019
Device Model NumberG058
Device Catalogue NumberG058
Device Lot Number171667
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/19/2022
Supplement Date Received by Manufacturer08/09/2022
Initial Report FDA Received Date06/08/2022
Supplement Report FDA Received Date08/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/06/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age56 YR