VYAIRE MEDICAL VELA VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number VELA
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/13/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Vyaire medical file identification: (b)(4).A vyaire field service representative(fsr) evaluated the device onsite and performed an initial check on the unit.O2 (oxygen) out of range occured.The fsr replaced the o2 cell and performed o2 calibrations.Verification check was performed and the unit is ready for use.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event or Problem Description
The customer reported to vyaire medical problems with vela ventilator where unit was alarming o2 range error due to the blender output being 8% out of specifications.At this time, there was no information regarding patient associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VELA VENTILATOR
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
VYAIRE MEDICAL
510 technology dr
irvine IL 92618
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
erika bonilla
510 technology dr
irvine, IL 92618
7149227837
MDR Report Key14640205
Report Number2021710-2022-16181
Device Sequence Number11988112
Product Code CBK
UDI-Device Identifier10846446001501
UDI-Public(01)10846446001501(11)20080205
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K093094
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVELA
Device Catalogue Number16531-00
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/13/2022
Initial Report FDA Received Date06/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/05/2008
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1