| Model Number |
1570-01-070 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Cyst(s) (1800); Foreign Body Reaction (1868); Necrosis (1971); Pain (1994); Distress (2329); Discomfort (2330); Deformity/ Disfigurement (2360); Osteolysis (2377); Metal Related Pathology (4530); Swelling/ Edema (4577); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
01/25/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Initial reporter occupation: lawyer.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Litigation complaint received ad 25 may 2022.After review of litigation records, plaintiff alleges severe pain and discomfort, increased metal levels in blood including cobalt and chromium; permanent injuries, emotional distress, disability, disfigurement and economic damages.Doi: (b)(6) 2007.Dor: (b)(6) 2022.Affected side: right hip.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide lot specific complaint database search, or manufacturing record evaluation, was not possible as the required lot number was not provided.The information received will be retained for trend analysis, post market surveillance, or other events within the quality system.Device history lot: a manufacturing record evaluation (mre), was not possible because the required lot code was not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.(removed the date of revision for the stem since this was not revised.).
|
| |
|
Event or Problem Description
|
|
Pfs provided product code of the asr component.However, there is no harm and product deficiency provided.After review of the medical records the patient was confirm revised to address elevated metal ions and local alval soft tissue reaction.Operative note reported turbid fluid approx.25-30 ml, necrotic tissue, caseous appearing tissues, erosive cystic changes consistent with the alval soft tissue reaction these were all removed.
|
| |
|
Event or Problem Description
|
|
Update ad 24 jun 2022.After review of medical records, it was reported that the patient was revised due to failed metal on metal right tha with elevated metal ions and local alval soft tissue reaction.Operative notes reported that upon incision a turbid fluid was encountered.The vast majority of the pericapsular tissue was necrotic.Upon inspecting the acetabulum, it was noted that there were at least 3 areas that demonstrated erosive cystic changes consistent with alval soft tissue reaction.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.(health effect - clinical code) previously reported appropriate term / code not available (e2402) used to capture the blood heavy metal increased is now updated to (e1618) for metal related pathology.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.A worldwide lot specific complaint database search, or manufacturing record evaluation, was not possible as the required lot number was not provided.The information received will be retained for trend analysis, post market surveillance, or other events within the quality system.Device history lot : a manufacturing record evaluation (mre), was not possible because the required lot code was not provided.
|
| |
|
Search Alerts/Recalls
|