| Model Number |
5803 |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Codes |
Fall (1848); Insufficient Information (4580)
|
| Date of Event |
05/10/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the table fell.
|
| |
|
Event or Problem Description
|
|
Table fall was reported with patient on.Customer also stated that the t-pin was missing but table still held in place until moved patient and then it dropped.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the information received from incident investigation, it was determined that there were no issues with the device and user error was contributed to falling of the table top with patient on.Inservice staff training was provided following the incident.
|
| |
|