Catalog Number PC803EU/A
Medical Device Problem Code Unraveled Material (1664)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/31/2022
Type of Reportable Event Serious Injury
Event or Problem Description
The account alleges that during an emergent pericardiocentesis procedure, the access guidewire coil started to unravel.The patient was hemodynamically unstable and diagnosed with cardiac tamponade at the time.The access wire was discarded, and a new kit was opened and used to complete this procedure for the patient.
 
Additional Manufacturer Narrative
The suspect device is not expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
 
Additional Manufacturer Narrative
The suspect device was not returned for evaluation.A photograph was provided by the account.The complaint is confirmed based on the photo received from the account.A root cause could not be conclusively determined.The device history record was reviewed, and no exception documents were found.A search of the complaint database was performed and no similar complaints for this lot number were identified.Should the device be returned later, the investigation will be re-opened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERICARDIOCENTESIS CATHETER
Common Device Name
IRRIGATION CATHETER
Manufacturer (Section D)
MERIT MEDICAL IRELAND LTD
parkmore business park west
ballybrit
galway,
EI 
Manufacturer (Section G)
MERIT MEDICAL IRELAND LTD
parkmore business park west
ballybrit
galway,
EI  
Manufacturer Contact
bryson heaton
1600 merit parkway
south jordan, UT 84095
MDR Report Key14641293
Report Number9616662-2022-00006
Device Sequence Number4867954
Product Code GBX
UDI-Device Identifier00884450013465
UDI-Public884450013465
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
N/A
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2024
Device Catalogue NumberPC803EU/A
Device Lot NumberK2319605
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/01/2022
Supplement Date Received by Manufacturer06/24/2022
Initial Report FDA Received Date06/09/2022
Supplement Report FDA Received Date06/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-