It was reported that nursing staff transferred a bariatric patient from a recliner to a bed using a maxi move lift.While the patient was over the bed, part of the sling material started to rip, and the sling detached from the lift.The staff was able to support the patient.It was reported that no injury was sustained due to the event.The nurse manager reported that the patient had a massive excess of skin and skin size centrally located in the lower abdominal area.After the event, the sling was picked up from the customer and returned for further evaluation to the manufacturer.During a sling inspection, the manufacturer noticed that sling material broke on the sling body area.The stitching of the sling was checked and no malfunction was found.It was indicated that the probable root cause of the reported issue can be associated with the device handling process performed by the caregivers.The manufacturer indicated 2 hypotheses of potential causes of reported sling ripped: - the patient weight mighte have not been evenly distributed in the sling.- the sling was caught by a sharp object.The clip sling instruction for use contains information that: ¿the patient¿ s psychical disabilities, weight distribution and general physique needs to be taken into consideration when selecting a sling.¿ post market surveillance data was reviewed for the last 3 years and no similar complaints weres found.To sum up, it was reported by the customer that the sling ripped during use, and from that perspective, the sling did not meet the manufacturer¿s specifications.No injury was reported.This complaint has been determined reportable since the sling ripped during use.
|