| Model Number |
M00565050 |
| Medical Device Problem Codes |
Difficult or Delayed Positioning (1157); Premature Activation (1484)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/13/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device was noted to be contaminated and will not be returned for analysis; therefore, a problem analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation on (b)(6) 2022 that a wallflex colonic stent was implanted in the colon to treat a 5cm intestinal obstruction during a stent placement procedure performed on (b)(6) 2022.The patient's anatomy was not tortuous prior to stent placement.During attempted deployment, the delivery shaft became stuck and the stent was unable to be deployed.Upon withdrawal of the delivery system, the stent deployed prematurely in an incorrect location.The stent remains implanted and another wallflex colonic stent was used to complete the procedure.There were no patient complications as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
| |
|