| Catalog Number |
0375562000 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/13/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the device broke during the procedure.The broken piece was able to be retrieved.
|
| |
|
Event or Problem Description
|
|
It was reported that the device broke during the procedure.The broken piece was able to be retrieved.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: the knife broke during a knee ligamentoplasty.The failure(s) identified in the investigation is consistent with the complaint record.The probable root cause/s could be the blade comes in contact with a hard object causing damage to the teeth and hitting the housing edges.Excessive pressure, such as bending or prying, may cause the arthroscopic shaver blades to bend/break.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
| |
|
Search Alerts/Recalls
|