|
(b)(4).Additional information was requested, and the following was obtained: did the outer package show any signs of damaged when received? no further information is available.Was the package properly sealed when received? no further information is available.Please provide the lot number: no further information is available.Device return status: we regularly contact with sales rep about the device returning.Lower left paper ¿ describing the specifications of the needle and suture was not attached and was not found in the plastic container.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.This is a combination product, and the event has been reviewed for both the suture and the triclosan.Trade name: irgacare, active ingredient(s): triclosan, dosage form: suture/solid/parenteral, strength: 275 g /m.Related to: 2210968-2022-04428.
|
|
Product complaint # (b)(4).Date sent to the fda: 6/29/2022.This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H3 investigational summary: one paper lid of product code d10083 was returned to ethicon for analysis.Upon visual analysis of the paper lid, was noted that the suture identification tab is not completely separated from the cover.Based on the information currently available, the suture identification tab not secure was identified during the investigation of the sample received.This product issue will be addressed through ethicon inc¿s quality system.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.Trade name - irgacare, active ingredient(s)- triclosan, dosage form - suture/solid/parenteral, strength -= 275 g /m.
|