| Model Number |
26183MC-S |
| Medical Device Problem Codes |
Break (1069); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/09/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During a laparoscopic supra cervical hysterectomy with bso, the loop snapped rather than cutting when cautery pedal was deployed.The surgeon completed the case with another small loop, with no impact to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The product has been returned, however, the evaluation has not been completed.
|
| |
|
Additional Manufacturer Narrative
|
|
Per evaluation from kst/mfr."based on the inspection of the instrument, the most probable root cause is that the customer operated the device incorrectly.Probably too much force was applied to the loop without activating the hf current, moreover, the current was too high, which led to overheating/damage to the instrument which has black discolorations at the breakage points.".
|
| |
|