• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARGON MEDICAL DEVICES JAWZ ENDOMYOCARDIAL BIOPSY FORCEPS 7F X 105CM STRAIGHT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARGON MEDICAL DEVICES JAWZ ENDOMYOCARDIAL BIOPSY FORCEPS 7F X 105CM STRAIGHT Back to Search Results
Model Number 190085
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/12/2022
Type of Reportable Event Malfunction
Event or Problem Description
The tissue cannot be taken out after using the product for the first time, and can only be taken out after replacing the same product for the second time.
 
Additional Manufacturer Narrative
The sample is indicated as available for return.As of the date of this report, the sample has not been returned.A follow-up report will be provided once the device has been received and reviewed.
 
Additional Manufacturer Narrative
The sample is indicated as returned.As of the date of this report, the sample has not been evaluated.A follow-up report will be provided once the device has been reviewed.
 
Event or Problem Description
The tissue cannot be taken out after using the product for the first time, and can only be taken out after replacing the same product for the second time.
 
Additional Manufacturer Narrative
A review of the dhr and inspection records was conducted, and no similar concerns were found.Ifu under precautions- "the forcep should be thoroughly rinsed with heparinized saline prior to each specimen collection during the procedure." one opened device was returned for evaluation.Device one would not actuate when received.Distal end of device was soaked for fifteen minutes in alcohol and removed.Device would then slightly start to open upon actuation.Jaw was manipulated open and then the device was soaked again and periodically actuated.After final removal from alcohol soak, the device would actuate.Root cause of the device not working properly is most likely a foreign substance(organic) caught in the clevis area mechanism of the distal tip(jawz).A result of the device not being thoroughly rinsed.Since this issue was believed to be caused by the user not thoroughly rinsing with heparinized saline, no corrective action is necessary.The ifu has been updated since the manufacture of this product to further stress the importance of cleaning the device upon removal.
 
Event or Problem Description
The tissue cannot be taken out after using the product for the first time, and can only be taken out after replacing the same product for the second time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JAWZ ENDOMYOCARDIAL BIOPSY FORCEPS 7F X 105CM STRAIGHT
Common Device Name
JAWZ ENDOMYOCARDIAL BIOPSY FORCEP
Manufacturer (Section D)
ARGON MEDICAL DEVICES
1445 flat creek rd
athens TX 75751
Manufacturer Contact
elnaz rahman
1445 flat creek rd
athens, TX 75751
9036759321
MDR Report Key14643514
Report Number0001625425-2022-01018
Device Sequence Number5195590
Product Code DWZ
UDI-Device Identifier00886333206834
UDI-Public00886333206834
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K951447
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date10/31/2025
Device Model Number190085
Device Catalogue Number190085
Device Lot Number5061276
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/31/2022
Supplement Date Received by Manufacturer05/31/2022
05/31/2022
Initial Report FDA Received Date06/09/2022
Supplement Report FDA Received Date07/18/2022
08/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient SexUnknown
-
-