• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ PANOPTIX TORIC UV ABSORBING TRIFOCAL IOL; LENS, MULTIFOCAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ PANOPTIX TORIC UV ABSORBING TRIFOCAL IOL; LENS, MULTIFOCAL Back to Search Results
Model Number TFAT40
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Blurred Vision (2137); Visual Disturbances (2140); Halo (2227)
Date of Event 10/19/2021
Type of Reportable Event Serious Injury
Event or Problem Description
A healthcare professional reported that following an intraocular lens (iol) implant procedure, the patient was unhappy with vision, patient stated that vision was blurry for distance and near, cannot see contrast and the vision at night was very bad with glare and halos.The iol was explanted in secondary procedure for different company lens.There was no patient harm, the surgeon prognosis was good, and patient was not hospitalized.No further information available.There are two medical device reports associated with this patient.This report is associated with the left eye.
 
Additional Manufacturer Narrative
The lens was returned inside a small plastic cup placed in a biohazard bag.Solution was dried on the lens.The optic was cut vertically into two pieces.The optic material splintered into cracks and splits along the line of damage.The optic edge of one optic portion was split.A power and resolution test could not be conducted on the returned lens due to being returned in pieces.Product history records were reviewed and the documentation indicated the product met release criteria.Associated product information was not provided.It is unknown if a qualified combination of products were used with the lens.The root cause cannot be determined for the reported complaint.The lens was returned cut into pieces with additional damage.Each lens was subjected to a 100% assessment of the power and optical resolution during the manufacturing process in order to determine acceptability per the lens model and diopter.The power and resolution of the returned lens could not be re-evaluated due to the optic damage.Information was provided in the file that the toric (2.25cyl add power +2.2/+3.2 ) 15.0 diopter lens was removed and replaced with a non-company lens (14.5 diopter).This information may suggest that the replacement lens model was an improved selection for the patient's vision needs or preferences.There are no other complaints in this lot.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ PANOPTIX TORIC UV ABSORBING TRIFOCAL IOL
Common Device Name
LENS, MULTIFOCAL
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14643743
Report Number1119421-2022-01230
Device Sequence Number5514411
Product Code MFK
UDI-Device Identifier00380652419130
UDI-Public00380652419130
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P040020
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/30/2024
Device Model NumberTFAT40
Device Catalogue NumberTFAT40.150
Device Lot Number12724974
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/31/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/16/2022
Initial Report FDA Received Date06/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
Patient SexFemale
-
-