| Model Number |
N/A |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/17/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The results of the investigation are as follows: review of the most recent repair record determined the fluid level was incorrect for both cylinders; not getting enough water due to leaking loop back check valve.The check valve was replaced and resolved the reported issue.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
It was reported that the machine reads not getting enough water, but it's full and water is on the floor.Loop back check valve needed to be replaced.Fluid volume was incorrect in both cylinders.There was no patient involvement as it did not occur during surgery.No adverse events were reported as a result of this malfunction.
|
| |
|