| Model Number |
1012632-59 |
| Medical Device Problem Code |
Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/18/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a moderately calcified vessel without tortuosity in the left iliac artery.The 9.0x59mm omnilink elite stent delivery system (sds) was inflated at 8 and 11 atmospheres (atms) and the stent did not come off the balloon and completely failed to deploy.The sds was removed and a 9.0x39mm omni elite sds was used and completed the procedure.There was no adverse patient effect and there was no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history identified no other incidents.The investigation was unable to determine a conclusive cause for the reported activation failure including expansion failures.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|
| |
|