H.6 investigation summary: scope of issue: the scope of issue is only limited to facscanto ivd 10 part # 657338, serial # (b)(6).Manufacturing defect trend: there are zero qns (quality notifications) related to the reported issue.Date range from 03jun2021 to date 03jun2022.Complaint trend: there are 4 complaints similar to this one of leakage of biohazard; pr# (b)(4).Date range from 03jun2021 to date 03jun2022.Manufacturing device history record (dhr) review: dhr part # 657338, serial # (b)(6).File # (b)(4) was reviewed.The instrument met all the manufacturing specifications prior to release.Investigation result / analysis: the investigation was performed and based on the review of the complaint trend, defect trend, dhr, risk analysis and servicemax, the root cause of a leak in the facscanto instrument was due to a blockage in the waste line.The user disconnected and connected the waste tank and performed startup.They removed the o-ring and the retainer, cleaned around and made sure that they were well set but this did not help resolve the issue.The technical support center specialist noted that the waste line must have been clogged.The fse confirmed the issue and cleared the tubing line by performing a flush to ensure proper flow and accuracy.No parts were requested for evaluation.After the repair the instrument was tested and was performing as expected.Although a leak was present, and no one was injured, users should wear proper ppe (personal protective equipment) while operating the instrument as stated in the bd facscanto¿ flow cytometer instructions for use (ifu), #23-14730-03 rev.1/vers.A, starting on page 147 in cleaning the surfaces.In addition, proper handling of fluidic filters can be found starting on page 149, maintenance.The safety risk of this hazard has been identified to be within the acceptable level.Service max review: review of related work order #: (b)(4), case #(b)(4) install date: (b)(6) 2014 defective part number: n/a work order notes: o subject / reported: 657338 - facscanto ivd 10 - vacuum error has been detected o problem description: during the cleaning an error message pops up.It looks like the aspirator arm is not acquiring the liquid.The tube is getting wet on the top.Error message: vacuum error has been detected.O work performed: flushed the waste lines using syringe ok.Measured vacuum at the waste pump ok.Run beads for 30 min, no error message.The instrument is ready for use.O cause: blockage o solution: flushed the waste lines returned sample evaluation: a return sample was not requested because no parts replaced.Risk analysis: risk management file part # 338942ra (revision 08, version g) facscanto product family risk analysis was reviewed.No new hazards have been identified and the current mitigations are sufficient.Hazard(s) identified? yes no o id: (b)(6) o hazard: potential leakage o cause: potential lack of workmanship during assembly o harmful effects: potential biohazard exposure o risk control: qualification and first article inspection o implementation verification: vp # 11392 protocol o effectiveness verification: vp # 11392 report o p: (b)(4) o s: (b)(4) o ri: (b)(4) o residual risk evaluation : afap o new hazards: none mitigation(s) sufficient yes no.Root cause: based on the investigation results the root cause of the leak was due to a blockage in the waste line conclusion: based on the investigation results, the root cause of the leak and why the aspirator arm could not acquire liquid was due to a blockage in the waste line.The field service engineer confirmed this issue when he flushed the waste line and the instrument was then ready to use.The safety risk of this hazard has been identified to be within the acceptable level.Supporting document: n/a.
|