• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC MICROPUNCTURE TRANSITIONLESS ACCESS SET; DYB INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC MICROPUNCTURE TRANSITIONLESS ACCESS SET; DYB INTRODUCER, CATHETER Back to Search Results
Model Number G43871
Medical Device Problem Code Unraveled Material (1664)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/24/2022
Type of Reportable Event Malfunction
Event or Problem Description
As reported, during a procedure involving placement of an arterial line, a micropuncture traditionless access set's wire "shredded" upon withdrawal of the wire.
 
Additional Manufacturer Narrative
It is unknown if the device will be returned.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Additional information: a2, a3, a4, b5, b7, d9, e4, h3, h6 (annexes e and f).This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
Additional information was received (b)(6) 022.Access was obtained in the right radial artery.The patient was reportedly hospitalized due to the disease process.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unchanged, or unavailable.Summary of event: as reported, during a procedure involving placement of an arterial line, a micropuncture transitionless access set's wire "shredded" upon withdrawal of the wire.Access was obtained in the right radial artery.The patient was reportedly hospitalized due to the disease process.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Investigation evaluation: reviews of the complaint history, device history record, instructions for use (ifu), manufacturing instructions, and quality control procedures were conducted during the investigation.The complaint device was not returned to cook for investigation.A document-based investigation evaluation was performed.One related non-conformance was found on a subassembly lot; however, affected product was scrapped and not replaced.There have been no other reported complaints for this lot number.Although the non-conformance on the subassembly lot is relevant to the reported failure mode, all non-conforming product was scrapped and there are 100% inspections to capture this non-conformance.Adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot related complaints have been received from the field; therefore, it was concluded that there is no evidence that non-conforming product exists in-house or in the field.The product ifu warns ¿advance the.018-inch wire guide through the introducer needle.Caution: do not withdraw the wire guide through the introducer needle; breakage may result.If the wire guide tip much be withdrawn while the needle is inserted, remove both the needle and the wire as a unit.¿ the information provided upon review of the dmr, dhr, and ifu suggests that there is evidence the device was manufactured to specification.There is no evidence of non-conforming devices in-house or in the field.Cook has concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.Based on the information provided and the results of the investigation, cook has concluded that a component failure unrelated to manufacturing or design deficiencies contributed to this incident.It is unknown if the wire was withdrawn through a needle.Details of the patient¿s anatomy are also unknown.The risk analysis for this failure mode was reviewed and no additional escalation was required.The appropriate personnel have been notified and cook will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information has been received since the last report was submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICROPUNCTURE TRANSITIONLESS ACCESS SET
Common Device Name
DYB INTRODUCER, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key14644810
Report Number1820334-2022-01029
Device Sequence Number17420536
Product Code DYB
UDI-Device Identifier00827002438712
UDI-Public(01)00827002438712(17)241213(10)14413212
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K171275
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/13/2024
Device Model NumberG43871
Device Catalogue NumberMPIS-501-NT-U-SST
Device Lot Number14413212
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/26/2022
Supplement Date Received by Manufacturer06/15/2022
08/29/2022
Initial Report FDA Received Date06/09/2022
Supplement Report FDA Received Date06/21/2022
09/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/13/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age55 YR
Patient SexMale
Patient Weight87 KG
-
-