• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS AG POLARCUP TRIAL INSERT NAV 22/59; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS AG POLARCUP TRIAL INSERT NAV 22/59; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 75023364
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/25/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, after a thr surgery, the plastic on the side of a polarcup trial insert nav 22/59 broke.Surgery was performed, without any delay, with the same device.As this happen after procedure patient was not harmed as consequence of this problem.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
H3, h6: it was reported that, after a total hip replacement surgery, the plastic on the side of a polarcup trial insert nav 22/59 broke.The device, used in treatment, was not returned for investigation.A product evaluation could therefore not be performed.A complaint history review was performed.The occurrence of the reported failure mode is within its expected risk level as per risk management.Furthermore, the concerning risk is covered through the smith + nephew risk management file.Due to insufficient information it is not possible to investigate whether the reported device met manufacturing specifications upon release for distribution and it is not possible to perform a review of past corrective actions.According to the document "processing (cleaning, disinfection and sterilization) of instruments from smith & nephew orthopaedics ag" (lit.N°03389-en 1363 v3 11/19), all instruments must be inspected and controlled for proper functioning after cleaning/disinfection.Based on available information, the reported failure mode could not be confirmed.A relationship between the reported event and the device cannot be confirmed.Due to insufficient information, it is not possible to speculate about factors which could have contributed to the reported event.No probable cause can be determined.The need for corrective action is not indicated.Nevertheless, smith + nephew will continue to monitor this device for similar issues.Should additional information become available, this complaint will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLARCUP TRIAL INSERT NAV 22/59
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ  CH-5000
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ   CH-5000
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14644827
Report Number9613369-2022-00272
Device Sequence Number5693130
Product Code LXH
UDI-Device Identifier07611996112535
UDI-Public07611996112535
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number75023364
Device Catalogue Number75023364
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/25/2022
Supplement Date Received by Manufacturer07/15/2022
Initial Report FDA Received Date06/09/2022
Supplement Report FDA Received Date07/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-